Model No: AGN-CM-192
It is designed for precise tissue sectioning at low temperatures. It is supplied with both high-profile and low-profile blade holders, along with a glass/Perspex anti-roll plate. The instrument features a Peltier-element cooling system that maintains a cooling temperature of −40 °C ± 5 °C for continuous operation, ensuring stable conditions and preventing unwanted freezing.

Share

Technical Specifications

  • Instrument type: Floor-standing cryostat with an encapsulated, splash-proof rotary microtome
  • Microtome type: Fully automatic microtome
  • Cooling system: Dual-compressor cooling combined with a Peltier system
  • Freezing shelf with storage capacity for approximately 10 ± 5 specimen chucks, featuring motorized coarse feed and manual sectioning capability
  • Microtome operation: Electronically controlled and motorized
  • Coarse feed speeds: Slow at 300 µm/sec (20 µm steps) and fast at 900 µm/sec
  • Dual-compressor feature ensures specimens remain frozen during cryo-bar (freezing station) power failure
  • Equipped with validated disinfection methods such as formaldehyde fumigation, UV, or ozone disinfection, effective against commonly known pathogens within the cryochamber
  • Encapsulated, splash-proof automated rotary microtome integrated into a floor-standing cryostat
  • Sectioning speeds: Slow 500–600 and fast 800–900
  • Blade holder with lateral movement and blade clamping/unclamping mechanism
  • Maximum specimen size: 55 × 55 mm or 40 × 55 mm
  • Section thickness range: 1 to 100 micrometers
  • Cryochamber fitted with bright white illumination and digital temperature display
  • Sectioning can be performed in both manual and fully motorized automatic modes
  • Flywheel/rotary wheel locking mechanism provided for enhanced user safety
  • Handwheel provision: One mechanical handwheel with electric brake
  • Motorized rapid and coarse feed speeds of 300 µm/sec (coarse) and 900 µm/sec (rapid)
  • Bright white chamber illumination
  • Trimming thickness range from 10 ± 5 micrometers up to 600 micrometers
  • Cooling temperature range: −45 ± 5 °C
  • Automatic and manual chamber defrost options, with at least one automatic defrost cycle every 24 hours and a defrost duration of 15 minutes or less
  • Section thickness settings include:
    • 0–5.0 µm in 0.5 µm steps
    • 0–20.0 µm in 1.0 µm steps
    • 0–60.0 µm in 5.0 µm steps
    • 0–100.0 µm in 10.0 µm steps
SALIENT FEATURES
  • Vertical stroke length: 60 mm ± 10 mm
  • Horizontal feed range: 25 ± 10 mm
  • Defrost cycle duration: Approximately 10 minutes
  • Power supply: 230 V, 50 Hz, single phase
  • Automatic specimen retraction provided on the return stroke